If Moku:Lab only has two output channels, how is it possible to generate feedback control signals as well as the modulation tone and scanning waveform?
-
Moku:Go
Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Lock-in Amplifier
- Moku:Lab
- Moku:Pro
The low-bandwidth control signal on output channel 2 can be separated electronically from the high-frequency modulation tone using an external bias-tee (not included with Moku:Lab). An appropriate bias-tee can be purchased from Mini-Circuits.
The scanning waveform is typically applied to the same actuator as the high-bandwidth control signal, so no bias-tee is required to separate the two signals.