Phasemeter output waveform resolution
Phasemeter output appears to have phase drift
-
Moku:Go
Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Lock-in Amplifier
- Moku:Lab
- Moku:Pro
The Moku:Lab Phasemeter instrument can generate an output waveform. This waveform is generated by a NCO ('numerically controlled oscillator') with an internal Moku:Lab reference of 10 MHz. Over a long period, it is possible to observe a steady slow phase drift when comparing this output waveform with another signal.
The precision of this output waveform is defined by the resolution of the NCO thus :
Any mismatch in the signals frequency leads to an observable but slow drift of phase.