How does Moku:Lab calculate the Power Spectral Density?
PSD and ASD calculation in the Phasemeter
-
Moku:Go
Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Lock-in Amplifier
- Moku:Lab
- Moku:Pro
In Moku:Lab's Phasemeter, Power Spectral Density (PSD) and the Amplitude Spectral Density (ASD) are calculated in the iPad App using Welch’s method of overlapping periodograms with a 50% overlap and a Hanning window. The number of points is either 512, 1024, or 2048 depending on the sampling rate chosen.